Analysis Instrument

Full Automatically Infrared Oil in Water Analyzer

Changsha Suny Electronic Technology Limited
  • Display:Digital
  • Appliance:Food Analysis, Biochemistry
  • Portable:Non Portable
  • Certification:CE, ISO9001
  • Type:Full Automatically Infrared Oil in Water Analyzer
  • Measurement Accuracy Error:<±2 %

Base Info

  • Model NO.:OT- EP3000R
  • Main Machine Net Weight:28kg
  • Main Machine External Dimension:450mm×350mm×380mm
  • Transport Package:Stand Export Package
  • Trademark:Labkits
  • Origin:Hunan China
  • Production Capacity:500PCS,Month

Description

Basic Info.

Model NO. OT- EP3000R Main Machine Net Weight 28kg Main Machine External Dimension 450mm×350mm×380mm Transport Package Stand Export Package Trademark Labkits Origin Hunan China Production Capacity 500PCS/Month

Product Description

Item: Full Automatically Infrared Oil in Water Analyzer
Model: OT- EP3000R
Full Automatically Infrared Oil in Water Analyzer

Summary:
OT- EP3000R type fully automatic oil analyzer is fully automatically infrared oil analyzer ,which is specially designed for water samples of high concentration, mainly used in the pollution sources, the measurement of oil content in industrial wastewater, suitable for petroleum and petrochemical steel company, sewage treatment plant, the monitoring of pollution source water samples
 
OT- EP3000R fully automatically infrared oil analyzer, without contacting with carbon tetrachloride, automatic filling, automatic extraction, automatic water and impurity removal, automatic measurement, automatic cleaning, automatic liquid discharge and store data.
 
Conform to national standard"HJ637-2012 water quality the determination of petroleum and animal and plant oil Infrared photometric method"
 
OT-EP3000R fully automatically infrared oil analyzer needn't contact with carbon tetrachloride. OT-EP3000R is composed of fully automatically operating software, infrared spectroscopic system, closed extraction system. Using extraction solvent (CCl4) in a certain proportion, extract oils in the water with the method of magnetism side stereo stir technology, then after removing water and impurity lead the extracted solution to cuvette by filter unit. Next infrared spectroscopic system measure and analyze. If installed with specialized magnesium silicate filter device, it could measure the content of petroleum and animal and plant oil. At last discharge waste and clean pipeline.
 
The equipment is mainly used in the measurement of oil content in the surface water, groundwater, industry waste water. It is applicable to oil detection in water in water supply company, hydrological stations, petroleum and petrochemical, steel smelting, environmental monitoring station, urban drainage units, etc.
 
Main instrument characteristics:
1. Full automation: Full automation, automatic sampling , extraction, dewater and filter, measurement, drainage, cleaning.
2. Healthy and safety: closed extraction system don't need the operator to participate in the process, not contacting with carbon tetrachloride.
3. The extracting method is accord with national standard HJ637-2012, the extraction result is consistent with GB Code Method.
4. Adopt imported filter device, instead of the function of anhydrous sodium sulfate, effectively removing  water and impurity;
5. Magnesium silicate adsorption device: realizing the automatic separation of mineral oil and the plants and animals, do not need adsorb and separate by manually;
6. The extraction system adopts completely anticorrosive and non-oleophylic material, run cleaning process, max reducing the cross contamination.
7. Spare parts quality assurance: adopt imported solenoid valve , peristaltic pump, air pump accessories ,
to ensure that the instruments are durable, effectively reducing the failure rate;
8. Manual and automatic switching, quick operation, easy to use.
 
Technical index:
Detection limit0.02mg/L (calculate 3 times standard deviation measuring 11 times blank)
Minimum detectable concentration0.001mg/L (enrichment extraction method)
Maximum measured concentration10000mg/L (dilution gaging)
Instrument measurement range(0.05~120)mg/L
RepeatabilityRSD1% (10-100mg/L Oil sample determined 11 times)
Measurement accuracy error<±2 %
Related coefficientR>0.999
sampling volume5ml--600ml 
single sample testing time5-12min
Main machine net weight28kg
Main machine external dimension450mm×350mm×380mm
 
Application unitsNational standard
The determination of environmental monitoring station, sewage and industrial wastewaterGB8978-1996 integrated wastewater discharge standard
Urban water supply company or running-water companyGB3838-2002 surface water quality standard
Petrochemical industriesGB3551-83 petroleum refining industrial water contaminant emission standard
hydrological environment monitoring stationGB3838-2002 surface water environmental quality standards
Railway environment monitoring stationGB5469-85 wagon Washing wastewater emission standard
thermal power plant environmental monitor stationGB8978-1996 integrated wastewater emission standard and other standard
smelting industry waste water monitoringGB13456-92 steel industrial water  contaminant emission standard
marine fishery water quality monitoringGB3097 Sea water quality standard and GB11607 water quality standard of fishery
Municipal drainage station or sewage
treatment plant
GB18918-2002 municipal sewage treatment plant contaminant emission standard
Agricultural products quality supervision and inspection center and Agricultural forestry environment monitoring stationGB4284-84 agricultural sludge pollution control standards