Analysis Instrument

Full Spectrum Direct Reading Spectrograph Spectrometer for Metal Analysis

Changsha Suny Electronic Technology Limited
  • Display:Digital
  • Appliance:Food Analysis, Biochemistry
  • Portable:Non Portable
  • Certification:CE, ISO9001
  • Type:Full Spectrum Direct Reading Spectrometer
  • Wavelength Range:130 ~ 800nm

Base Info

  • Model NO.:ICP-TY-9000
  • Operating Temperature:(10 ~ 35)°c
  • Storage Temperature:(0 ~ 45)°c
  • Transport Package:Stand Export Package
  • Trademark:Labkits
  • Origin:Hunan China
  • Production Capacity:500PCS,Month

Description

Basic Info.

Model NO. ICP-TY-9000 Operating Temperature (10 ~ 35)°c Storage Temperature (0 ~ 45)°c Transport Package Stand Export Package Trademark Labkits Origin Hunan China Production Capacity 500PCS/Month

Product Description

Item: Full Spectrum Direct Reading Spectrograph Spectrometer for Metal Analysis For Sale 
Model: ICP-TY-9000
Full Spectrum Direct Reading Spectrograph Spectrometer for Metal Analysis
Specification and technical parameter
ItemIndicator
Detection substrateFe
Detection timeAccording to the type of test sample, usually it is about 25s.
Optical systemPalmer type - runge Rowland circle full spectrum vacuum type optical system
Wavelength range130 ~ 800nm
Focal length400mm
DetectorHigh-performance CCD array
ElectrodeTungsten material spray electrode
Analytical gapSample Stage Analytical gap:3.4mm
Operating temperature(10 ~ 35)°C
Storage temperature(0 ~ 45)°C
Operating humidity20~ 80%
Argon purity requirements99.999%
Argon gas inlet pressure0.5MPa
Argon flowStimulating flow is about 3.5L/min, maintaining flow is about 0.4L/min.
Stimulating max power400VA
Light source typebrand new adjustable digital light source, high energy preliminary combustion science (HEPS)
Discharge frequency100-1000 Hz
Discharge currentMax 400A
Spark Stand aperture13mm
Vacuum systemVacuum software automatically control and monitor
Working power220V AC, 50/60Hz,  the single-phase power supply of protective grounding
Instrument dimensions860*720*500
Instrument weight80kg
 
Optical system
l full spectrum optical system of Palmer type - runge structure
l  Maximum wavelength range: 130 ~ 800nm
l Multiple high performance CCD detector
l Integrated optical chamber process molding, resistant to change of environment temperature
l Vacuum light chamber design guarantees C, S, P, could achieve the best performance.
l light room temperature control, constant temperature is 35 °C
l Directed type optical technology and lens MgF2 materials ensure that the best energy of UV wavelength of C, S, P.