Biostellar High Precision Gas Chromatograph for Chemical Laboratory
Prosperity Biotech (Shandong) Co., Ltd.- Appliance:The Analysis of Environmental Samples
- Ray Type:Gamma Ray Spectrometer
- Gaspath System:Double
- Detector:MSD
- Energy Resolution:≤8%
- Certification:CE
Base Info
- Model NO.:BS-GC112A
- Spectrometer Detector:SDD
- Stationary Phase:Liquid Solvent
- Mobile Phase:Gas
- Chromatography Type:Gas Chromatography
- Chromatographic Column Type:Convetional Analytical Column
- Chromatographic Column Length:10~20cm
- Chromatographic Column Diameter:0.2~0.5mm
- Transport Package:Carton or Wooden Package
- Specification:2-10ml
- Trademark:BIOSTELLAR
- Origin:China
- Production Capacity:500000units,Month
Description
Model
BS-GC112A
Oven
Capacity
22L
Temp. Range
RT+5ºC~400ºC
Temp.Accuracy
±0.1ºC
Temp.Program
9 step
Program Total Time
9999.9 min
MaxTemp. Heating Rate
60ºC/min
Sample Injector
Temp.Range
RT+7ºC~420ºC
Column or split injector can be installed.
Carrier gas flow control type: Constant pressure mode.
Max. Quantity
3pcs
Pre Column Pressure Range
0-400kpa
Pre Column Pressure Accuracy
0.1kpa
Flow Setting Range
H2: 0-200ml/min;
N2: 0-150ml/min
Detector
Temp.Range
RT+7ºC~420ºC
Hydrogen Flame Ionization Detector (FID)
Logarithmic amplifier
High voltage switch control
Baseline signal display
Detection limit:≤3*10-12g/s
(Sample: n-hexadecane)
Baseline noise:≤5*10-14A
Baseline draft:≤6*10-13A
Dynamic range: 107
RSD:≤3%
Thermal Conductivity Detector (TCD)
(Optional)
Sensitivity: 8000mv.ml/mg
(Sample: n-hexadecane)
Baseline noise:≤0.05mV
Baseline draft:≤0.15mV/30min
Dynamic range: 105
Power Supply
AC220V±10%, 50Hz
Standard Accessory
FID detector,Capillary injection system, Chromatography workstation software,RS232 port, LAN port
Optional Accessory
TCD detector, Chromatographic column, Gas purifier.
Package Size (W*D*H)
740*730*840mm
Gross Weight
80kg