Analysis Instrument

Sy-B037 Electrophoresis Machine Automatic Electrophoresis Analyzer for Lab

Sunny Medical Equipment Limited
  • Sample Type:Solidity
  • Processing Method:Mix
  • Warranty:1 Year
  • Environment:Normal Pressure and Temperature
  • Display:Digital Display
  • Performance:Semi-automatic

Base Info

  • Model NO.:SY-B037
  • Power Supply:220V
  • Weight:0-50kg
  • Certification:Meds
  • Customized:Customized
  • Name:Electrophoresis Analyzer
  • Current Range:0~400mA
  • Voltage Shift:≤0.2%
  • Voltage Range:0~600V
  • Transport Package:Carton
  • Trademark:SUNNYMED
  • Origin:China
  • Production Capacity:800 Sets,Month

Description

Voltage range 0~600VCurrent range 0~400mAVoltage shift ≤0.2%Current shift ≤0.2%Working mode   Constant current or constant voltageOutput format   in four groups synchronouslyPower supply 220V±10%  50HzFuse BGXP-5  5ASy-B037 Electrophoresis Machine Automatic Electrophoresis Analyzer for Lab
Sy-B037 Electrophoresis Machine Automatic Electrophoresis Analyzer for Lab
Sy-B037 Electrophoresis Machine Automatic Electrophoresis Analyzer for LabSy-B037 Electrophoresis Machine Automatic Electrophoresis Analyzer for Lab

Operating instructions

 

1) Connect the output ends to the gel box. The red end is positive electrode (anode) and the black one is 

negative electrode (cathode).

 

2) Revolve the knob of "Adjust" anticlockwise until it stops.

 

3) Press the button of "Power" to switch on power supply and lighten the digital panel. 

 

4) Choose a constant mode

 

constant voltage mode

 

   Press the button of "constant T" to turn on the indicator light of "constant T".

 

   Now the electrophoresis apparatus is working on constant voltage mode.

 

constant current mode

 

   Press the button of "constant A" to turn on the indicator light of "constant A".

 

   Now the electrophoresis apparatus is working on constant current mode.

 

5) Revolve the knob of "Adjust" clockwise gently until it reaches a certain value of voltage or current.

 

Cautions

 

1) DO NOT connect the electrophoresis apparatus to the gel box with the power on. Also, DO NOT pull out 

the output cables that connecting the apparatus to the gel box with the power on. Always Connect the 

electrophoresis apparatus to the gel box BEFORE turning the power supply ON.

 

2) Assume that you want to set the voltage value to "150V" at the constant voltage mode. If the current value 

becomes extremely high or even overflows the rated value (400mA) before you could adjust the voltage value 

to 150V, do examine whether the electrophoresis apparatus is well connected to the gel box or the 

electrophoresis buffers is accurately prepared. 

 

3) When more than one gel boxes are working together, choose the constant voltage mode to reduce the 

interactions between boxes.

 

4) PULL the power cables OUT OF the power supply when changing the fuse. Specification of the new fuse 

should accord with that indicated in "Technical data"