  • Accuracy:0.2% or 0.5%
  • Primary Current Output:1000A
  • Voltage Output:2500V
  • CT Ratio:25000A,5A(5000A,1A)
  • PT Ratio:1-500kv
  • Weight:23kg

Base Info

  • Model NO.:GF106T
  • Standards:IEC60044-1, IEC60044-6, IEC61869-2-2012
  • Transport Package:Standard Export Carton
  • Specification:CE, ISO9001: 2008, IEC
  • Trademark:GFUVE
  • Origin:Beijing, China
  • Production Capacity:200set,Month

Description

9. Actual secondary load test10. Resistance test
PARAMETERS
Electrical parametersElectrical parameters - continuedStandardsMechanical parametersEnvironmental conditionsAccuracy0.2%, 0.5%Power supplyAC 220V±10% or AC 110V±10%, 50/60HzExcitation voltage output0-2500Vrms, Excitation current output0-5Arms (20A peak-value)Large current output0-1000AAccuracy≤0.5% (0.2%RD+0.3%RG)Secondary winding DC resistance measurementRange0.1-300ΩAccuracy≤0.5% (0.2%RD+0.3%RG)Secondary actual load measurementRange5VA-1000VAAccuracy≤0.5% (0.2%RD+0.3%RG)±0.1VACT/PT phase error measurementAccuracy±4minResolution0.01minCT ratio error measurementRange≤25000A/5A(5000A/1A)Accuracy≤0.5%PT ratio error measurementRange1-500KVAccuracy≤0.5%Reference standardsGB1207-2006, GB1208-2006, GB16847-1997 
IEC60044-1, IEC60044-6, IEC61869-2-2012Safety standardsGB 4793.1-2007EMCEMC standard 89/336/EECFCC Subpart B of Part 15 Class AIEC 1000-4-2 /3 /4 /6Overall dimension (L x W x H) (mm)400 x 250 x 250 Weight (kg)  ≤22Relative humidity   90RH%Operating temperature-10°C to +40°CAltitude≤1000m