Analysis Instrument

Conductivity Controller Power Water Plant

Shanghai Boqu Instrument Co., Ltd.
  • After-sales Service:Provided
  • Warranty:Provided
  • Type:Conductivity Meter
  • Analysis Method:Chemical Analysis
  • Display:LED
  • Application:Water Treatment

Base Info

  • Model NO.:DDG-3080
  • Portable:Non-portable
  • Physical Analysis Type:Distilling Layered Method
  • Chemical Analysis Type:Karl Fischer Method
  • Operating Mode:Keyboard
  • Off Mode:Manual
  • Transport Package:Cartonbox
  • Specification:144*144*104mm
  • Trademark:BOQU
  • Origin:China
  • HS Code:9027809900
  • Production Capacity:5000,Year

Description

Basic Info.

Model NO. DDG-3080 Portable Non-portable Physical Analysis Type Distilling Layered Method Chemical Analysis Type Karl Fischer Method Operating Mode Keyboard Off Mode Manual Transport Package Cartonbox Specification 144*144*104mm Trademark BOQU Origin China HS Code 9027809900 Production Capacity 5000/Year

Product Description

Conductivity Controller Power Water Plant

DDG-3080 Industrial Conductivity Meter is our company's latest top-class instrument equipped with a microprocessor and complete English menu. It has display and menu operation pattern; it is characterized by its intelligent operation, multi-function and high measuring performance and its strong environmental adaptability. It can be widely used for continuous monitoring of the conductivities of solutions used in the following fields: thermal power plant, chemical fertilizer, chemical industry, metallurgy, environmental protection, pharmacy manufacturing, biochemistry, foodstuff and running water.

Features
It has a complete English display and friendly interface. Various parameters can be displayed at the same
time: conductivity, output current, temperature, time and status. Bitmap type liquid crystal display module
with high resolution is adopted. All the data, status and operation prompts are displayed in English. There
is no symbol or code that is defined by the manufacturer.

Technical Indexes



 
1.Conductivity measuring range:  0.01~20μS/cm  (Electrode: K=0.01)
                          0.1~200μS/cm  (Electrode: K=0.1)
                            1.0~2000μS/cm (Electrode: K=1.0)
                            10~20000μS/cm (Electrode: K=10.0)
                           30~600.0mS/cm  (Electrode: K=30.0)
2. Intrinsic error of the electronic unit: conductivity: ±0.5%FS, temperature: ±0.3ºC
3. Range of automatic temperature compensation: 0~199.9ºC, with 25ºC as the reference temperature
4. Water sample tested: 0~199.9ºC, 0.6MPa
5. Intrinsic error of the instrument: conductivity: ±1.0%FS, temperature: ±0.5ºC
6. Automatic temperature compensation error of the electronic unit: ±0.5%FS
7. Repeatability error of the electronic unit: ±0.2%FS±1 Unit
8. Stability of the electronic unit: ±0.2%FS±1 unit/24h
9. Isolated current output: 0~10mA ( load<1.5kΩ)
                     4~20mA (load<750Ω)  (double-current output for option)
10. Output current error: ≤±l%FS
11. Error of electronic unit caused by ambient temperature: ≤±0.5%FS
12. Error of the electronic unit caused by the supply voltage: ≤±0.3%FS
13. Alarm relay: AC 220V, 3A
14. Communication interface: RS485 or 232 (optional)
15. Power supply: AC 220V±22V, 50Hz±1Hz, 24VDC (optional)
16. Protection grade: IP54 , aluminium shell suitable for outdoor use
17. Clock accuracy: ±1 minute/month
18. Data storage capacity: l month (1 point/5 minutes)
19. Saving time of data under a continuous power-failure condition: 10 years
20. Overall dimension: 146 (length) x 146 (width) x 150 (depth) mm; dimension of the hole: 138 x 138mm
21. Working conditions: ambient temperature: 0~60ºC; relative humidity <85%
22. Weight: 1.5kg
23. The conductivity electrodes with the following five constants are usable: K=0.01, 0.1, 1.0, 10.0, and 30.0.
Conductivity Controller Power Water PlantConductivity Controller Power Water PlantConductivity Controller Power Water PlantConductivity Controller Power Water PlantConductivity Controller Power Water PlantConductivity Controller Power Water PlantConductivity Controller Power Water Plant