Analysis Instrument

Laser Diffraction Particle Size Analyzers (Bettersize2000LD)

Ningbo Yinzhou Hybers Imp.& Exp. Co., Ltd.
  • Size Range:0.02-2000 Um(Wet Mode); 0.1-2000 Um(Dry Mode)
  • Transport Package:Fumitation Wooden Box
  • Specification:CE
  • Trademark:BETTERSIZE
  • Origin:Dandong, China

Base Info

  • Model NO.:Bettersize2000LD
  • HS Code:90314990
  • Production Capacity:1000sets,Year

Description

Size Range : 0.02-2000 um(wet mode); 0.1-2000 um(dry mode)
Sample Inflow Method:  Circulating pump(wet mode); Turbulent flow(dry mode)
Built-in automatic circulating and dispersing system
Repeatability (wet mode): Less than 1% (CRM D50)
Accuracy (wet mode): Less than 1% (CRM D50)
Repeatability (dry mode): Less than 3% (CRM D50)
Accuracy (dry mode): Less than 3% (CRM D50)
Measuring Principle: Mie scattering theory
Laser: high quality optical fiber semiconductor laser
Operation System: Win XP/ Win 7
Connection Port : USB or RS232
Photodetector: 90pcs
Testing Time: 2-3 minutes per test
Ultrasonic Power: 50-100 W
Tank Capacity: 500 ml
Circulating Flowrate: 500-2500 ml / min
Voltage: 220V, 50/60 Hz
Pressure: 0.1-0.8Mpa, continuously adjustable
Gas flow: 400-6000L/min
Outer Dimension: 800*500*418mm
Net weight: 64kg