Analysis Instrument

Qualitative and Quantitative SF6 Gas Leak Detector (NDIR)

HV Hipot Electric Co., Ltd.
  • Brand:HV Hipot
  • Customized:Customized
  • Portable:Portable
  • Gas:Sf6 etc.
  • Type:Electrochemical Gas Analyzer
  • Measurement Range:0-1000ppmv

Base Info

  • Model NO.:GDWG-IV
  • Storage Temperature:-20~+60 centigrade
  • Operation Temperature:-20~+50 centigrade
  • Working Voltage:220VAC±10%,50Hz Or Built-in Li-battery
  • Sensitivity:0.1ppmv (can be customized to 0.01ppm, for option)
  • Resolution:0.1ppmv
  • Response time:≤3s
  • Dimension:140*235*83mm
  • Weight:About 1.5kg
  • Transport Package:Wooden Case
  • Specification:ISO, SGS
  • Trademark:HV Hipot
  • Origin:Wuhan China
  • HS Code:9027100090
  • Production Capacity:2000PCS,Year

Description

Measuring principle

Non-dispersive infrared sensor (NDIR)

Resolution

0.1ppmv

Measuring range

0-1000ppmv

Accuracy

±2%FS (0-1000ppm). Even the measuring range is over 5000ppm, the instrument will not be damaged.

Repeat-ability error

≤±1%

Sensitivity

0.1ppmv (can be customized to 0.01ppm, for option)

Response time

3s

Recovery time

5s

Stability

≤±20ppm, more than 1000hours

Zero shift

≤±1%(F.S/year)

Linear error

≤±1%

Storage temperature

-20~+60ºC

Operation temperature

-20~+50ºC

Environment humidity

0-95%RH

Operation humidity

0-95%(non-condensing)

Sampling mode

Built-in electromagnetic diaphragm pump, automatic slot-in type, flow up to 1L/min.

Working voltage

220VAC±10%,50Hz Or Built-in Li-battery, continuous working 12hours after full charge.

Peak current

<700mA

Average power

<2W.

Operation air pressure

800~1150hPa.

Dimension

140*235*83mm

Weight

About 1.5kg