Tpc-3000 Fully Automatic Flash Point Analyzer (Closed-Cup)
Chongqing TOP Oil Purifier Co., Ltd.- After-sales Service:Online Service
- Warranty:1 Year
- Display:Color Touch LCD Display Screen
- Test Method:Thermocouple Differential Detection
- Self-Checking:Automatic Fault Checking
- Power Supply:AC220V
Base Info
- Model NO.:TPC-3000
- Dimension:397*305*240mm
- Transport Package:International Standard Export Box
- Specification:CE
- Trademark:TOP TESTER
- Origin:China
- HS Code:9031809090
- Production Capacity:100 Sets,Month
Description
Basic Info.
Model NO. TPC-3000 Dimension 397*305*240mm Transport Package International Standard Export Box Specification CE Trademark TOP TESTER Origin China HS Code 9031809090 Production Capacity 100 Sets/MonthProduct Description
TPC-3000 Fully Automatic Flash Point Analyzer (Closed-Cup)
TPC-3000 fully automatic flash point analyzer(closed-up) is designed in accordance with ASTM D93 IGB261-2008, SH/T0315 etc. It adopts microcomputer technology to achieve completely automatic operation. It has the function of self fault checking ,highly accurate testing, with strong repeatability, stable and reliable performance. It is an ideal flash point analyzer(closed-up) for petroleum,chemical,power station,environmental,railway,scientific research etc.
Main Features:
1.Color touch LCD screen display, man-machine conversation interface in English with menu instructions.
2.Adopts high speed signal processor: precision simulation/digital converter circuit.
3.Adopts PID control algorithm: automatically adjust the heating curve according to standards.
4.Built-in automatic atmospheric pressure detector: automatic to detect and correct the atmosphere pressure and self-adjust the flash point.
5.Aspiration,stirring,testing,alerting,cooling,printing are fully automatic.
6.Automatically finish the whole testing process, opening the lid, alerting, cooling, printing etc.
7.Security airless source of platinum alloy :stable,reliable and high precision
8.Differential detection: automatic correction of systematic bias; automatic stop testing by over temperature detection and alert.
9.With parameters reminders like: test date, test time etc.
10.Data storage: store 300 groups testing data.
Technical parameter
Display | Color touch LCD display screen |
Test method | Thermocouple differential detection |
Test range | Room temperature-350ºC |
Resolution | 0.1ºC |
Repeatability | National standard: flash point<104ºC,allowed ±1ºC; flash point≥104ºC allowed ±2ºC National standard: flash point≤110ºC,allowed ±1ºC; flash point>110ºC allowed ±2ºC |
Printer | High speed thermal printer,paper size:55mm |
Temperature test | Imported platinum resistance (Pt100Ω) |
Ignition mode | Automatically electronic without need of gas source |
Self-checking | automatic fault checking |
Power | <550w |
Temperature | Atmospheric temperature: 10ºC-45ºC,humidity: 30-80% |
Power supply | AC220V±10% |
Dimension | 397*305*240 mm |
Weight | About 15 kg |