Analysis Instrument

Tpc-3000 Fully Automatic Flash Point Analyzer (Closed-Cup)

Chongqing TOP Oil Purifier Co., Ltd.
  • After-sales Service:Online Service
  • Warranty:1 Year
  • Display:Color Touch LCD Display Screen
  • Test Method:Thermocouple Differential Detection
  • Self-Checking:Automatic Fault Checking
  • Power Supply:AC220V

Base Info

  • Model NO.:TPC-3000
  • Dimension:397*305*240mm
  • Transport Package:International Standard Export Box
  • Specification:CE
  • Trademark:TOP TESTER
  • Origin:China
  • HS Code:9031809090
  • Production Capacity:100 Sets,Month

Description

Basic Info.

Model NO. TPC-3000 Dimension 397*305*240mm Transport Package International Standard Export Box Specification CE Trademark TOP TESTER Origin China HS Code 9031809090 Production Capacity 100 Sets/Month

Product Description

TPC-3000 Fully Automatic Flash Point Analyzer (Closed-Cup)
Tpc-3000 Fully Automatic Flash Point Analyzer (Closed-Cup)
Application:
TPC-3000 fully automatic flash point analyzer(closed-up) is designed in accordance with ASTM D93 IGB261-2008, SH/T0315 etc. It adopts microcomputer technology to achieve completely automatic operation. It has the function of self fault checking ,highly accurate testing, with strong repeatability, stable and reliable performance. It is an ideal flash point analyzer(closed-up) for petroleum,chemical,power station,environmental,railway,scientific research etc.

Main Features:
1.Color touch LCD screen display, man-machine conversation interface in English with menu instructions.
2.Adopts high speed signal processor: precision simulation/digital converter circuit.
3.Adopts PID control algorithm: automatically adjust the heating curve according to standards.
4.Built-in automatic atmospheric pressure detector: automatic to detect and correct the atmosphere pressure and self-adjust the flash point.
5.Aspiration,stirring,testing,alerting,cooling,printing are fully automatic.
6.Automatically finish the whole testing process, opening the lid, alerting, cooling, printing etc.
7.Security airless source of platinum alloy :stable,reliable and high precision
8.Differential detection: automatic correction of systematic bias; automatic stop testing by over temperature detection and alert.
9.With parameters reminders like: test date, test time etc.
10.Data storage: store 300 groups testing data.

Technical parameter
Display Color touch LCD display screen
Test methodThermocouple differential detection
Test rangeRoom temperature-350ºC
Resolution0.1ºC
RepeatabilityNational standard: flash point<104ºC,allowed ±1ºC; flash point≥104ºC allowed ±2ºC
National standard: flash point≤110ºC,allowed ±1ºC; flash point>110ºC allowed ±2ºC
PrinterHigh speed thermal printer,paper size:55mm
Temperature testImported platinum resistance (Pt100Ω)
Ignition modeAutomatically electronic without need of gas source
Self-checking automatic fault checking
Power<550w
TemperatureAtmospheric temperature: 10ºC-45ºC,humidity: 30-80%
Power supplyAC220V±10%
Dimension397*305*240 mm 
WeightAbout 15 kg