Karl Fischer Water Content Tester (TP-2100)
Chongqing TOP Oil Purifier Co., Ltd.- After-sales Service:Online Service
- Warranty:1 Year
- Type:Karl Fischer Moisture Meter
- Display:LED
- Portable:Non-portable
- Operating Mode:Touch Screen
Base Info
- Model NO.:TP-2100
- Off Mode:Automatic
- Method of Display:7 Inch Color Touch LCD with English Display
- Method of Testing:Mpu Controlled Electrolysis(Karl Fischer Coulometr
- Rate of Electrolysis:2.4mgh2o,Min
- Electrolysis Current:Auto-Control Within 430mA
- Testing Range:0.01ugh2o-200mgh2o
- Sensitivity:0.01ugh2o
- Drift Compensation:Auto-Control Via Mpu
- Stirring Speed:Adjustable and Steerable
- Data Input:Input Via Touch Screen
- Transport Package:International Standard Export Box
- Specification:CE
- Trademark:TOP TESTER
- Origin:China
- HS Code:9031809090
- Production Capacity:60 Sets,Month
Description
Basic Info.
Model NO. TP-2100 Off Mode Automatic Method of Display 7 Inch Color Touch LCD with English Display Method of Testing Mpu Controlled Electrolysis(Karl Fischer Coulometr Rate of Electrolysis 2.4mgh2o/Min Electrolysis Current Auto-Control Within 430mA Testing Range 0.01ugh2o-200mgh2o Sensitivity 0.01ugh2o Drift Compensation Auto-Control Via Mpu Stirring Speed Adjustable and Steerable Data Input Input Via Touch Screen Transport Package International Standard Export Box Specification CE Trademark TOP TESTER Origin China HS Code 9031809090 Production Capacity 60 Sets/MonthProduct Description
Karl Fischer Water Content Tester (TP-2100)Introduction
TP-2100 water content tester adopts the global recognized classical moisture testing method - Karl Fischer Coulometric Method which is the most accurate method with low cost. With the universal standard and analyzing method, it can quickly analyze the moisture content in liquid, solid and gas and directly display testing results, with testing range from 0.0001%(1ppm) to 100%. Therefore, this testing method has been widely used in institutions and enterprises of petroleum, chemical industry, power, pharmaceuticals, pesticide, railways, research, etc. And it meets ASTM D6304, ISO 60814 etc.
The whole testing process is automatically controlled by this tester itself. Its auto-stirring and analyzing would be finished within 60seconds, and results would be displayed directly. Totally closed titration system avoids damage to operator from reagent.
Application Area
Liquid | Chemical Industry:alcohols, ethers, esters, alkenes, benzenes, amines, organic solvents, phenols and other organic products applicable to Karl Fischer Method for determination Petroleum and Power Industry:Insulation Oil, Transformer Oil, Turbine Oil, etc. Pharmaceutical Industry :medicine materials Pesticide Field:emulsifier Other Industry:electrolyte of lithium battery |
Solid | Various mineral salt, citric acid, explosives, paraffin and other solids with good solubility |
Gas | Natural gas, liquefied gas, Freon, butadiene, chloromethane, etc. |
Technical Parameters
Method of Display | 7 inch color touch LCD with English display |
Method of Testing | MPU controlled electrolysis(Karl Fischer Coulometric Method) |
Rate of Electrolysis | 2.4mgH2O/min |
Electrolysis Current | Auto-control within 430Ma |
Testing Range | 0.01ugH2O-200mgH2O |
Water Content Range | 0.0001%(1ppm)-100% |
Sensitivity | 0.01ugH2O |
Accuracy | Water Content: 2ugH2O-100ugH2O; Deviations≤±1ug, Water Content: 100ugH2O-1000ugH2O; Deviations≤±2.9ug, Water Content: >1000ugH2O; Deviations≤±0.2%(excluding deviations of sample injection) |
Drift Compensation | Auto-control via MPU |
Stirring Speed | Adjustable and steerable |
End-point Indication | LCD display /output via printing / voice warning / end-point indication light |
Data Input | Input via touch screen |
Sample Number | Numbered by users |
Storage | Infinite storage capacity |
Printer | Thermal printer, high-speed, low-consumption, with 44 millimeter-wide paper |
Printing Content | ug / ppm / mg/L / sample number /experimenter / date |
Calendar/Clock | automatic display and print analyzing time and date |
Self-checking | Breakdown self-checking |
Special Function | Work independently or connected with laptop or other computers |
Power Consumption | <40W |
Dimension | 390×270×190mm |
Working Environment | Temperature: 5ºC-40ºC,Humidity: <85% |
Power Supply | AC220V±10% 50HZ±2.5HZ |
Weight | About 7kg |